The solution in the small compartment A is added to the solution in the large compartment B after breaking the frangible pin or the peel seal immediately before use. The reconstituted solution shall be clear and colourless.
A package leaflet with detailed instruction for use is enclosed in the box.
Aseptic technique shall be used throughout the handling and administration to the patient.
Use only if the overwrap is undamaged, all seals are intact, frangible pin or peel seal is not broken, and the solution is clear. Press bag firmly to test for any leakage. If leakage is discovered, discard the solution immediately since sterility can no longer be assured.
The large compartment B is fitted with an injection port for the possible addition of other necessary drugs after reconstitution of the solution. It is the responsibility of the user to judge the compatibility of an additive medication with Phoxilium by checking for eventual colour change and/or eventual precipitation, insoluble complexes or crystals. Before adding a medication, verify if it is soluble and stable in this medicine and that the pH range of Phoxilium is appropriate (pH of reconstituted solution is 7.0–8.5). Additives may be incompatible. The Instructions for Use of the medication to be added must be consulted.
Remove any fluid from the injection port, hold the bag upside down, insert the drug through the injection port and mix thoroughly. The introduction and mixing of additives must always be performed prior to connecting the solution bag to the extracorporeal circuit. The solution must be administered immediately.
If a frangible pin separates the two compartments of the bag and a frangible pin is located in the luer connector the following instructions for use shall be followed:
I Remove the over wrap from the bag immediately before use and discard any other packaging materials. Open the seal by breaking the frangible pin between the two compartments of the bag. The frangible pin will remain in the bag.
II Make sure all the fluid from the small compartment A is transferred into the large compartment B.
III Rinse the small compartment A twice by pressing the mixed solution back into the small compartment A and then back into the large compartment B.
IV When the small compartment A is empty: shake the large compartment B so that the contents mix completely. The solution is now ready for use and the bag can be hung on the equipment.
V The dialysis or replacement line may be connected to either of the two access ports.
Va If the luer access is used, using aseptic technique, remove the cap and connect the male luer lock on the dialysis or replacement line to the female luer receptor on the bag: tighten. Using both hands, break the blue frangible pin at its base, and move it back and forth. Do not use a tool. Verify that the pin is completely separated and that the fluid is flowing freely. The pin will remain in the luer port during the treatment.
Vb If the injection port is used, first remove the snap-off cap. The injection port is a swabbable port. Then introduce the spike through the rubber septum. Verify that the fluid is flowing freely.
If a frangible pin separates the two compartments of the bag and a valve is located in the luer connector the following instructions for use shall be followed:
I Remove the over wrap from the bag immediately before use and discard any other packaging materials. Open the seal by breaking the frangible pin between the two compartments of the bag. The frangible pin will remain in the bag.
II Make sure all the fluid from the small compartment A is transferred into the large compartment B.
III Rinse the small compartment A twice by pressing the mixed solution back into the small compartment A and then back into the large compartment B.
IV When the small compartment A is empty: shake the large compartment B so that the contents mix completely. The solution is now ready for use and the bag can be hung on the equipment.
V The dialysis or replacement line may be connected to either of the two access ports.
Va If the luer access is used, remove the cap with a twist and pull motion, and connect the male luer lock on the dialysis or replacement line to the female luer receptor on the bag using a push and twist motion. Ensure that the connection is fully seated and tighten. The connector is now open. Verify that the fluid is flowing freely.
When the dialysate or replacement line is disconnected from the luer connector, the connector will close and the flow of the solution will stop. The luer port is a needle-less and swabbable port.
Vb If the injection port is used, first remove the snap-off cap. The injection port is swabbable port. Then introduce the spike through the rubber septum. Verify that the fluid is flowing freely.
If a peel seal separates the two compartments of the bag and a valve is located in the luer connector the following instructions for use shall be followed:
I Remove the over wrap from the bag immediately before use and discard any other packaging materials. Open the seal by holding the small compartment with both hands and squeezing it until an opening is created in the peel seal between the two compartments.
II Push with both hands on the large compartment until the peel seal between the two compartments is entirely open.
III Secure complete mixing of the solution by shaking the bag gently. The solution is now ready for use, and can be hung on the equipment.
IV The dialysis or replacement line may be connected to either of the two access ports.
IVa If the luer access is used, remove the cap with a twist and pull motion, and connect the male luer lock on the dialysis or replacement line to the female luer receptor on the bag using a push and twist motion. Ensure that the connection is fully seated and tighten. The connector is now open. Verify that the fluid is flowing freely.
When the dialysis or replacement line is disconnected from the luer connector, the connector will close and the flow of the solution will stop. The luer port is a needle-less and swabbable port.
IVb If the injection port is used, first remove the snap-off cap.The injection port is a swabbable port. Then introduce the spike through the rubber septum. Verify that the fluid is flowing freely
The reconstituted solution shall be used immediately. If not used immediately, the reconstituted solution should be used within 24 hours including the duration of the treatment after addition of the solution A to solution B.
The reconstituted solution is for single use only.
Discard any unused solution immediately after use.
Any unused medicinal product or waste material should be disposed of in accordance with local requirements.