09/01/2024
11. DOSIMETRY
The data listed in Table 1 are from ICRP (International Commission on Radiological Protection) Publication 128.
Table 1: Absorbed radiation doses for rubidium-82 (Rb-82) chloride
| Organ | Dose absorbed per unit activity administered (mGy/MBq) |
| | Adults | 15 years old | 10 years old | 5 years old | 1 year old |
| Adrenals | 2.4 x 10-3 | 3.6 x 10-3 | 5.1 x 10-3 | 7.0 x 10-3 | 1.0 x 10-2 |
| Bone surfaces | 4.2 x 10-4 | 5.6 x 10-4 | 8.5 x 10-4 | 1.4 x 10-3 | 3.1 x 10-3 |
| Brain | 1.4 x 10-4 | 1.4 x 10-4 | 1.6 x 10-4 | 1.9 x 10-4 | 2.8 x 10-4 |
| Breast | 1.9 x 10-4 | 2.0 x 10-4 | 1.3 x 10-2 | 2.2 x 10-2 | 4.3 x 10-2 |
| Gallbladder | 7.2 x 10-4 | 8.5 x 10-4 | 1.2 x 10-3 | 2.0 x 10-3 | 5.7 x 10-3 |
| Stomach | 8.3 x 10-4 | 1.1 x 10-3 | 1.6 x 10-3 | 2.7 x 10-3 | 5.4 x 10-3 |
| Small intestine | 2.0 x 10-3 | 2.6 x 10-3 | 4.6 x 10-3 | 7.7 x 10-3 | 1.5 x 10-2 |
| Colon | 1.1 x 10-3 | 1.4 x 10-3 | 2.5 x 10-3 | 4.1 x 10-3 | 7.8 x 10-3 |
| Ascending colon | 1.1 x 10-3 | 1.4 x 10-3 | 2.5 x 10-3 | 4.1 x 10-3 | 7.9 x 10-3 |
| Descending colon | 1.1 x 10-3 | 1.4 x 10-3 | 2.4 x 10-3 | 3.9 x 10-3 | 7.6 x 10-3 |
| Heart | 4.0 x 10-3 | 5.2 x 10-3 | 8.2 x 10-3 | 1.3 x 10-2 | 2.4 x 10-2 |
| Kidneys | 9.3 x 10-3 | 1.1 x 10-2 | 1.6 x 10-2 | 2.4 x 10-2 | 4.3 x 10-2 |
| Liver | 9.8 x 10-4 | 1.3 x 10-3 | 2.0 x 10-3 | 3.0 x 10-3 | 5.8 x 10-3 |
| Lungs | 2.6 x 10-3 | 3.8 x 10-3 | 5.5 x 10-3 | 8.5 x 10-3 | 1.7 x 10-2 |
| Muscles | 2.3 x 10-4 | 3.6 x 10-4 | 7.2 x 10-4 | 2.2 x 10-3 | 4.3 x 10-3 |
| Oesophagus | 1.5 x 10-3 | 2.4 x 10-3 | 3.7 x 10-3 | 8.1 x 10-3 | 1.5 x 10-2 |
| Ovaries | 5.0 x 10-4 | 4.9 x 10-4 | 1.2 x 10-3 | 2.0 x 10-3 | 4.4 x 10-3 |
| Pancreas | 2.6 x 10-3 | 3.7 x 10-3 | 7.6 x 10-3 | 9.7 x 10-3 | 2.1 x 10-2 |
| Red bone marrow | 3.8 x 10-4 | 4.6 x 10-4 | 7.8 x 10-4 | 1.5 x 10-3 | 3.8 x 10-3 |
| Skin | 1.8 x 10-4 | 2.3 x 10-4 | 3.7 x 10-4 | 6.1 x 10-4 | 1.2 x 10-3 |
| Spleen | 1.8 x 10-4 | 3.9 x 10-4 | 2.4 x 10-3 | 2.8 x 10-3 | 3.8 x 10-3 |
| Testicles | 2.6 x 10-4 | 3.3 x 10-4 | 5.0 x 10-4 | 7.9 x 10-4 | 1.5 x 10-3 |
| Thymus | 1.5 x 10-3 | 2.4 x 10-3 | 3.7 x 10-3 | 8.1 x 10-3 | 1.5 x 10-2 |
| Thyroid | 3.1 x 10-4 | 3.8 x 10-4 | 6.2 x 10-4 | 1.0 x 10-3 | 1.9 x 10-3 |
| Urinary bladder wall | 1.8 x 10-4 | 3.9 x 10-4 | 2.4 x 10-3 | 2.8 x 10-3 | 3.8 x 10-3 |
| Uterus | 1.0 x 10-3 | 1.1 x 10-3 | 1.5 x 10-2 | 2.3 x 10-2 | 4.1 x 10-2 |
| Other tissue | 3.1 x 10-4 | 5.0 x 10-4 | 9.3 x 10-4 | 2.1 x 10-3 | 4.7 x 10-3 |
| Effective dose per activity administered (mSv/MBq) | 1.1 x 10-3 | 1.4 x 10-3 | 3.0 x 10-3 | 4.9 x 10-3 | 8.5 x 10-3 |
For rubidium-82 (Rb-82), the effective dose resulting from administration of a maximal activity of 2220 MBq is 2.44 mSv.
For this activity of 2200 MBq, the typical radiation doses delivered to the critical organs below are as follows: kidneys: 20.65 mGy, heart: 8.88 mGy, lungs: 5.77 mGy and pancreas: 5.77 mGy.
12. INSTRUCTIONS FOR PREPARATION OF RADIOPHARMACEUTICALS
12.1 Operator training
The user must have undergone training by the generator distributor and must be given specific documentation. When the product is first introduced to a nuclear medicine site, all personnel must undergo training under the supervision of the Marketing Authorisation Holder.
Cardiogen-82 (Rb-82 generator) must only be used with an appropriate infusion system designed specifically for use with the Cardiogen-82 generator e.g. Cardiogen-82 Infusion System Model 510 or Model 1701.
12.2 Operating principle:
A syringe pump is used to pass sterile NaCl 9 mg/mL solution through the generator.
Only a NaCl 9 mg/mL solution that meets pharmacopeia quality standards and with no impurities or additives should be used for generator elution. The use of any other solutions (in particular those containing calcium, even in trace amounts, or additives) is strictly forbidden, as this may result in significant breakthrough of the parent nuclide Sr-82 with potential adverse consequences for the patient.
The eluate obtained from the generator contains rubidium-82. If the activity of the eluate as measured by a positron detector reaches a sufficient level, the valve directs the eluate toward the patient infusion line. The desired activity level and therefore the onset of infusion are pre-defined electronically.
12.3 Radionuclidic purity and quality control: measurement of Rb-82, Sr-82 and Sr-85 concentrations
At the start of each day the generator is used, before administering the product to the first patient, it is essential that the radionuclidic purity of the eluate be tested as follows:
The first eluate of the day must be discarded to ensure that a false positive for strontium breakthrough does not occur. This first 50 mL of eluate should be discarded (the Model 1701 infusion system will do this automatically), with due consideration of proper safety precautions. After regeneration of the column (10 minutes), daily quality control testing (rubidium (Rb-82) activity and levels of Sr-82 and Sr-85 in the eluate) should be performed carefully on the fixed 50 ml volume eluted for the second elution and with strict adherence to the instructions below. Records of test results should be kept.
Information on eluting the Cardiogen-82 generator follows below:
- waterproof gloves and effective protection should be worn while handling the rubidium (Rb-82) chloride solution.
- aseptic techniques should be employed throughout the preparation and elution processes.
- leave at least 10 minutes between each elution so as to allow for regeneration of the Rb-82.
- elute only with Sodium Chloride Infusion 9 mg/mL (0.9% w/v) which meets pharmacopeia quality standards and contains no impurities or additives.
- traceability in terms of infusion volume and activity should be ensured for the entire procedure.
12.4 Limits
The generator must not be used if any of its expiration limits are reached:
- 42 days have passed since the calibration date,
- a total elution volume of 17 L has already passed through the column since first use of the generator,
- eluate Sr-82 level exceeds 1x10-5 MBq/MBq Rb-82 after elution,
- eluate Sr-85 level exceeds 1x10-4 MBq/MBq Rb-82 after elution.
12.5 Additional tests
It is also necessary to carry out an additional quality control test if one of the following alert limits is reached:
- a total elution volume of 14 L has passed through the column since first use of the generator,
- eluate Sr-82 level exceeds 2x10-6 MBq/MBq Rb-82 after elution,
- eluate Sr-85 level exceeds 2x10-5 MBq/MBq Rb-82 after elution.
These additional quality controls should also be performed at times determined by the day's elution volume. Tests should be performed every 750 mL.
- For example, if the clinical site has eluted less than 750 mL from the generator during the day, no additional test is performed that day
- If the same clinical site elutes 1500 mL from the generator the next day, it will have to perform 3 tests that day:
1) the routine test required prior to use on the first patient,
2) a test after 750 mL has been eluted,
3) a test after 1500 mL has been eluted.
As soon as an alert limit is reached and throughout the life of the generator, additional quality tests should be performed after 750 mL has been eluted, in accordance with the procedures described below.
12.6 Procedure and calculation methods
Use Cardiogen-82 with an appropriate infusion system designed specifically for use with the Cardiogen-82 generator e.g. Cardiogen-82 Infusion System Model 510 or Model 1701.
- If using the Cardiogen-82 Infusion System Model 510, refer to the Eluate Testing Protocol in Section 12.6.1
- If using the Cardiogen-82 Infusion System Model 1701, refer to the Eluate Testing Protocol in Section 12.6.2
Follow instructions in the Cardiogen-82 Infusion System Model 510 or Model 1701 Operator's Manual for the set up and intravenous infusion of rubidium (Rb 82) chloride injection dose(s).
12.6.1 Cardiogen-82 Infusion System Model 510 Eluate Testing Protocol
Levels of rubidium Rb-82, Sr-82 and Sr-85 are determined using an ionisation chamber-type dose calibrator.
Procedures 1 to 11 below should be performed.
The rubidium (Rb-82) chloride content of the injectable solution is determined as follows:
1. Set a dose calibrator for Rb-82 as recommended by the manufacturer, or use the Co-60 setting and divide the reading obtained by 0.548. Read off the value given by the instrument in MBq (megabecquerel).
2. Elute the generator aseptically using 50 mL of additive-free Sodium Chloride Infusion 9 mg/mL (0.9% w/v) of pharmacopeia quality, and discard the eluate (first elution).
3. Allow at least 10 minutes for the regeneration of Rb-82 and then elute the generator aseptically with 50 mL of sodium chloride 9 mg/mL solution for injection (pharmacopeia quality, with no additives or impurities) at a rate of 50 mL/minute and collect the eluate in a stoppered glass vial (plastic containers should not be used). Note the exact time (hh:mm:ss) of the end of elution (EOE).
4. Using the dose calibrator, determine the activity of the Rb-82 (ARb (t)) and note the time (t) of the reading. Correct the reading to the end of elution using the relevant fraction remaining for Rb-82 (see Table 5).
ARb (EOE) = ARb (t) / Fraction Remaining
Example: if the reading is taken 2.5 minutes after the end of elution, correction for decay is carried out by dividing the dose calibrator reading by 0.25.
To measure the concentration of Sr-82 in the eluate, proceed as follows:
5. Using the sample obtained for determination of Rb-82 activity, let the sample stand for at least one hour to allow Rb-82 to decay completely.
6. Measure the activity of the sample using a dose calibrator at the setting recommended by the manufacturer for Rb-82 and/or Sr-82. Alternative method: use the Co-60 setting and divide the reading obtained by 0.548. Read off the value given by the instrument in MBq (megabecquerel).
7. Read off the ratio of Sr-85/Sr-82 on the calibration date from the generator label. Use Table 2 to read off the correction factor for the ratio of Sr-85/Sr-82 on the day of use after calibration.
Calculate the ratio R using the following formula:
on calibration date x correction factor for the ratio on the date of measurement
8. Apply a correction factor (F) of 0.478 to compensate for the contribution of Sr-85 to the measured value.
9. Calculate the amount of Sr-82 in the sample using the following equation:
| Sr-82 (MBq) = | measured value (MBq) [1 + (R x F)] |
Example:
| | Dose calibration reading (MBq) (=measured value) = 2.96 x 10-2 Sr-85/Sr-82 ratio on day of calibration: 1.0172 Day of use after day of calibration: 22 Correction factor for the ratio at 22 days according to Table 2 1.46 R = 1.0172 x 1.46 = 1.48 Correction factor (F) = 0.478 Sr-82(MBq) = 2.96 x 10-2/[1 + (1.48 x 0.478)] Sr-82(MBq) = 1.734 x10-2 |
10. Determine the Sr-82 content by dividing the MBq of Sr-82 by the MBq of Rb-82 at the end of elution.
Example:
| | 1.734 x10-2 MBq of Sr-82 1850 MBq of Rb-82 at the end of elution (1.734 x10-2 MBq Sr-82)/(1850 MBq Rb-82) = 9.4 x 10-6 MBq/MBq Rb-82 |
In this example, the Sr-82 content is greater than the alert limit of 2x10-6 MBq/MBq of Rb-82. Consequently, an additional breakthrough test should be performed.
11. Determine the Sr-85 content by multiplying the result obtained in step 10 by the ratio (R) Sr-85/Sr-82.
Example:
| | 9.4 x 10-6 x 1.48 = 1.4 x 10-5 MBq Sr-85/MBq Rb-82 |
In this example, the Sr-85 content is less than the alert limit of 2x10-5 MBq/MBq of Rb-82.
Table 2: Sr-85/Sr-82 ratio
| Days | Correction factor for the ratio | Days | Correction factor for the ratio |
| 0* | 1.00 | 22 | 1.46 |
| 1 | 1.02 | 23 | 1.48 |
| 2 | 1.03 | 24 | 1.51 |
| 3 | 1.05 | 25 | 1.53 |
| 4 | 1.07 | 26 | 1.56 |
| 5 | 1.09 | 27 | 1.59 |
| 6 | 1.11 | 28 | 1.61 |
| 7 | 1.13 | 29 | 1.64 |
| 8 | 1.15 | 30 | 1.67 |
| 9 | 1.17 | 31 | 1.70 |
| 10 | 1.19 | 32 | 1.73 |
| 11 | 1.21 | 33 | 1.76 |
| 12 | 1.23 | 34 | 1.79 |
| 13 | 1.25 | 35 | 1.82 |
| 14 | 1.27 | 36 | 1.85 |
| 15 | 1.29 | 37 | 1.88 |
| 16 | 1.31 | 38 | 1.91 |
| 17 | 1.34 | 39 | 1.95 |
| 18 | 1.36 | 40 | 1.98 |
| 19 | 1.38 | 41 | 2.01 |
| 20 | 1.41 | 42 | 2.05 |
| 21 | 1.43 | | |
* day of calibration
Radiation emission
The half-value layer is 0.7 cm of lead (Pb). Table 3 shows a range of values for the relative attenuation of the radiation emitted by radionuclides, which results from the interposition of various thicknesses of lead. For example, the use of a 7.0 cm thickness of lead will attenuate the radiation emitted by a factor of about 1000.
Table 3: Radiation attenuation by lead shielding
| Shield thickness (Pb, in cm) | Attenuation factor |
| 0.7 | 0.5 |
| 2.3 | 10-1 |
| 4.7 | 10-2 |
| 7.0 | 10-3 |
| 9.3 | 10-4 |
Strontium-82, with its half-life of 25 days (600 hours), decays to rubidium-82 (Rb-82). To correct for physical decay of strontium Sr-82, Table 4 shows the fractions that remain at selected intervals after the time of calibration.
Table 4: Radioactive decay table:
| Days | Fraction remaining | Days | Fraction remaining | Days | Fraction remaining |
| 0* | 1.000 | 15 | 0.660 | 30 | 0.435 |
| 1 | 0.973 | 16 | 0.642 | 31 | 0.423 |
| 2 | 0.946 | 17 | 0.624 | 32 | 0.412 |
| 3 | 0.920 | 18 | 0.607 | 33 | 0.401 |
| 4 | 0.895 | 19 | 0.591 | 34 | 0.390 |
| 5 | 0.871 | 20 | 0.574 | 35 | 0.379 |
| 6 | 0.847 | 21 | 0.559 | 36 | 0.369 |
| 7 | 0.824 | 22 | 0.543 | 37 | 0.359 |
| 8 | 0.801 | 23 | 0.529 | 38 | 0.349 |
| 9 | 0.779 | 24 | 0.514 | 39 | 0.339 |
| 10 | 0.758 | 25 | 0.500 | 40 | 0.330 |
| 11 | 0.737 | 26 | 0.486 | 41 | 0.321 |
| 12 | 0.717 | 27 | 0.473 | 42 | 0.312 |
| 13 | 0.697 | 28 | 0.460 | | |
| 14 | 0.678 | 29 | 0.448 | | |
* Calibration date
To correct for physical decay of rubidium (Rb-820, Table 5 shows the fraction of the rubidium (Rb 82) chloride injectable solution remaining in 15-second intervals up to 300 seconds after time of calibration.
Table 5: Radioactive decay table:
| Seconds | Fraction remaining | Seconds | Fraction remaining |
| 0* | 1.000 | 165 | 0.218 |
| 15 | 0.871 | 180 | 0.190 |
| 30 | 0.758 | 195 | 0.165 |
| 45 | 0.660 | 210 | 0.144 |
| 60 | 0.574 | 225 | 0.125 |
| 75 | 0.500 | 240 | 0.109 |
| 90 | 0.435 | 255 | 0.095 |
| 105 | 0.379 | 270 | 0.083 |
| 120 | 0.330 | 285 | 0.072 |
| 135 | 0.287 | 300 | 0.063 |
| 150 | 0.250 | | |
*Time elapsed since elution.
12.6.2 Cardiogen-82 Infusion System Model 1701 Eluate Testing Protocol
The rubidium (Rb-82) chloride content of the injectable solution is determined as follows:
1. Set a dose calibrator for Rb-82 as recommended by the manufacturer. Obtain the reading from the instrument in millicuries.
2. Following the prompts in the Graphical User Interface (GUI) for the Cardiogen-82 Infusion System Model 1701, elute the generator with additive-free Sodium Chloride Infusion 9 mg/mL (0.9% w/v) at a rate of 50 mL/min and collect the eluate in the stoppered vial specifically provided for use with the Cardiogen-82 Infusion System Model 1701 (alternative vials, glass or plastic are not suitable). Note the exact time of end of elution (EOE).
3. Using the external dose calibrator, assay the eluate at exactly 2:30, 3:45, or 5:00 minutes after EOE.
4. Following the prompts in the GUI for the Cardiogen-82 Infusion System Model 1701, enter the Rb-82 reading from the dose calibrator and the time since EOE.
5. The infusion system software will automatically calculate the Calibration Ratio.
- If the ratio is within +/- 2% (0.98 to 1.02), the infusion system will allow acceptance of the calibration factor that was used for the elution.
- If the ratio is not within +/- 2% (0.98 to 1.02), the system requires another calibration elution (steps 1 through 4).
6. Repeat steps 1 through 4 for a flow rate of 20 mL/min.
Perform additional system calibration every 14 days.
To measure the concentration of Sr-82 in the eluate, proceed as follows:
Each day, before administering rubidium Rb-82 chloride injection, perform the following test, including Mandatory Eluate Testing:
1. Place the stoppered vial, which is specifically provided for use with the Cardiogen-82 Infusion System, Model 1701 (alternative vials, glass or plastic are not suitable) in the Sr detector well on the Cardiogen-82 Infusion System Model 1701 and, following the prompts in the GUI for the infusion system, initiate the Daily Quality Control workflow.
2. The infusion system will automatically perform the Sr Detector Background Reading.
3. The infusion system will automatically perform the Generator Column Wash.
4. Strontium Level Test and Dose Constancy:
a. The infusion system will elute the generator with 50 mL of additive-free 0.9% Sodium Chloride Injection USP at a rate of 50 mL/min into the stoppered vial (which is specifically provided for use with the Cardiogen-82 Infusion System Model 1701).
b. The Sr detector measures the Rb-82 and strontium in the 50 mL elution.
c. The infusion system software will automatically calculate the Sr-82 and Sr-85 levels on the day (post calibration) of the measurement using the ratio of Sr-85/Sr-82 on the day of calibration provided on the generator label, and using the full exponential decay calculation for each, accounting for the generator's age.
d. Using the Rb-82 and strontium measurements, the infusion system software will automatically calculate MBq Sr-82/MBq Rb-82 and MBq Sr-85/MBq Rb-82. The GUI will automatically indicate if the results exceed Alert or Expiration Limits.
e. The infusion system software will automatically calculate Dose Constancy.
5. Constancy Check of the Sr detector: The infusion system GUI will prompt the user to perform the constancy check of the Sr detector.
a. Place the external constancy source in the detector well of the infusion system.
b. The infusion system software will automatically calculate the constancy of the Sr detector versus the external constancy source when instructed.