Allow ELIGARD 22.5 mg to come to room temperature by removing from the refrigerator not less than 30 minutes prior to use. Eligard can remain out of the fridge for up to 4 weeks at temperatures below 25 degrees.
Please prepare the patient for injection first, followed by the preparation of the product, using the instructions below. If Eligard is not prepared using the proper technique, it should not be administered, as lack of clinical efficacy may occur due to incorrect reconstitution of the product.
Step 1
On a clean field, open the tray by tearing off the foil from the corners to remove the contents. Discard the desiccant pouch. Remove the pre-connected syringe system (Figure 1.1) from the tray. Open the safety needle package, available under the tray (Figure 1.2) by peeling back the paper tab. The enclosed safety needle must be used only. Note: Syringe A and Syringe B should not be lined-up yet.
| Figure 1.1 Tray Contents: pre-connected syringe system | Figure 1.2 Under the Tray: safety needle and cap |
Step 2
Press the latching button on the connector with your finger and thumb (Figure 2) until you hear a snapping sound (“Click”). The two syringes will be lined up. No particular orientation of the syringe system is required to activate the connector. Do not bend the syringe system (please note that this may cause leakage as you may partially unscrew the syringes).
Step 3
Hold the syringes horizontally, gently push the liquid contents of Syringe A into the leuprorelin acetate powder contained in Syringe B. Thoroughly mix the contents by gently pushing the plungers back and forth between the syringes for 60 cycles until a homogenous, viscous solution is obtained (a cycle is one push of the plunger for Syringe A and one push of the plunger for Syringe B) (Figure 3). Do not push both plungers at the same time as it may cause leakage. Do not bend the syringe system (please note that this may cause leakage as you may partially unscrew the syringes).

When thoroughly mixed, the viscous solution will appear with a colour in the range of colourless to shades of white or pale yellow.
Important: After mixing proceed with the next step immediately as the product gets more viscous over time. Do not refrigerate the mixed product.
Please note: Product must be mixed as described; shaking WILL NOT provide adequate mixing of the product.
Step 4
After mixing, hold the syringes vertically with Syringe B on the bottom. The syringes should remain securely coupled. Draw the entire mixed product into Syringe B (short, wide syringe) by pushing down the Syringe A plunger and slightly withdrawing the Syringe B plunger (Figure 4).
Step 5
While ensuring Syringe A plunger is fully pushed down, hold the connector and unscrew it from Syringe B. Syringe A will remain attached to the connector (Figure 5). Ensure that no product leaks out as the needle will then not secure properly when attached.
Please note: one large or a few small air bubbles may remain in the formulation - this is acceptable. Please do not purge the air bubbles from Syringe B as product may be lost!
Step 6
• Hold Syringe B upright and hold back the white plunger to prevent loss of the product.
• Secure the safety needle to Syringe B by holding the syringe and gently turning the needle clockwise with approximately a three-quarter turn until the needle is secure (Figure 6).
Do not over tighten as this may cause cracking of the needle hub resulting in leakage of the product during injection. The safety shield may also be damaged if the needle is screwed with too much force.
Should the needle hub crack, appear to be damaged, or have any leakage, the product should not be used. The damaged needle should not be substituted/replaced and the product should not be injected. The entire product should be disposed of securely
In the event of damage to the needle hub, use another ELIGARD kit.
Step 7
Move the safety shield away from the needle and pull off the protective needle cover immediately prior to administration (Figure 7).
Important: Do not operate the safety needle mechanism before administration. Should the needle hub appear to be damaged, or leak, the product should NOT be used. The damaged needle should NOT be replaced and the product should NOT be injected. In the event of damage to the needle hub, use another ELIGARD kit.
Step 8
It is not necessary to purge the air bubbles as it poses no risks for the subcutaneous injection. Purging air bubbles might cause product loss. Inject the product subcutaneously whilst keeping the safety shield away from the needle.
Administration Procedure:
| • Select an injection site on the abdomen, upper buttocks, or another location with adequate amounts of subcutaneous tissue that does not have excessive pigment, nodules, lesions, or hair and has not recently been used. • Cleanse the injection-site area with an alcohol swab (not enclosed). • Pinch the skin around the injection site with your thumb and forefinger. • Quickly insert the needle at a 90° angle into the skin. Release the skin. • Inject the entire contents of Syringe B using a slow, steady push. Ensure that the full amount of the product in Syringe B is injected before removing the needle. • Withdraw the needle quickly at the same 90° angle used for insertion while maintaining pressure on the plunger. | |
Step 9
After injection, lock the safety shield using any of the activation methods listed below.
a. Closure on a flat surface
Press the safety shield, lever side down, onto a flat surface (Figure 9a) to cover the needle and lock the shield.
Verify locked position through audible and tactile “click”. Locked position will completely cover needle tip.
b. Closure with your thumb
Placing your thumb on the safety shield (Figure 9b), cover the needle tip and lock the shield.
Verify locked position through audible and tactile “click”. Locked position will completely cover needle tip.
Once safety shield is locked, immediately dispose of the needle and syringe in an approved sharps container.